Testing Products for ASTM F382
APP-F382
Application Insight:
Bio-Simulation: Bone plates must survive millions of cycles of walking loads without snapping.
Compliant Hardware Solution
313-50-1200
This machine has been validated to meet the specific requirements of this standard.
View Machine DetailsRegulatory & Test Details
| App Id | APP-F382 |
|---|---|
| App Standard | ASTM F382 |
| Standard Organization | ASTM |
| Test Type | Flexure |
| Material Type | Medical |
| Specimen Type | Bone Plate |
| Industry Type | Medical / Ortho |
| Min Force Kn | 5 |
| Max Force Kn | 50 |
| Min Spd Mm/m | 0.1 |
| Max Spd Mm/m | 5 |
| Expert Spd | 0.1mm for static, Hz for fatigue. |
| Mach Recommend | 313-50-1200 |
| Pointer App Software | N-ASTM-F382 |
| Pointer Fixture Family | G-ASTM-F382 |
| Pointer Strain Family | E-ASTM-F382 |
Implementation Commentary
This 4-point bend test is the primary FDA-clearance hurdle for any new orthopedic implant design.